Analysis of high-precision vanadium isotope ratios by medium resolution MC-ICP-MS

Thumbnail Image
Date
2015-11
Authors
Nielsen, Sune G.
Owens, Jeremy D.
Horner, Tristan J.
Linked Authors
Alternative Title
Date Created
Location
DOI
Related Materials
Replaces
Replaced By
Keywords
Abstract
We present and verify a new method to measure vanadium isotope ratios using a Thermo Scientific Neptune multi-collector inductively-coupled plasma mass spectrometer (MCICP- MS) operated in medium mass resolution mode. We collect masses 48 through 53 simultaneously using the L2, L1, Center, H1, H2 and H3 collectors. The Center cup is equipped with a 1012 ; resistor, H1 is equipped with a 1010 ; resistor, while the rest of the collectors have standard 1011 ; resistors. Unlike previous low-resolution methods, the use of medium mass resolution (=M/M ~ 4,000) permits separation of V, Ti and Cr isotopes from all interfering molecular species representing combinations of C, N, O, S, Cl, and Ar. We show that the external reproducibility follows a power law function with respect to the number of V+ ions collected and achieve an external reproducibility of ± 0.15 ‰ with total V+ ion beam intensities of ~ 1 nA. The separation of interfering molecular species from the V mass spectrum reduces the V requirement for precise isotope data to as little as 200-300 ng V per analysis — a reduction of ~90% compared with previous methods — making several low-V matrices amenable to V isotope analysis.
Description
Author Posting. © The Author(s), 2015. This is the author's version of the work. It is posted here by permission of Royal Society of Chemistry for personal use, not for redistribution. The definitive version was published in Journal of Analytical Atomic Spectrometry 31 (2016): 531-536 , doi:10.1039/C5JA00397K.
Embargo Date
Citation
Cruises
Cruise ID
Cruise DOI
Vessel Name